Avox 1000e manual
Forgot your password? First time users, please register here. At present, Werfen operates directly in over 30 countries and in more than territories through distributors. Hong Kong. Werfen Hong Kong.
South Korea. Czech Republic. Germany Nicolai. Hungary Comesa. Poland Comesa. Tatlisu Mah,Sanli Cad. United Kingdom. North America. United Arab Emirates. Toggle navigation. About Werfen. Learn more. Enter the new lot number, using the number keys, then press.
Press followed by to confirm the new lot number. Note: Lot numbers of liquid controls must be entered before they can be tagged to a liquid QC test see page Insert a cuvette filled with liquid quality control material see page A prompt to confirm liquid QC is displayed: 2.
A prompt to confirm that the level and lot numbers are being entered is displayed: 3. A menu for selection of a control level is displayed: 4. Select the level that is being run and press. A menu for selection of a lot number is displayed: 5. Select the previously entered lot number that is being run and press. The sample results will be displayed on the screen, and the lot number and LQC level will be added to the record for that sample.
Press followed by to display the current QC Lockout setting: 3. A screen is displayed for specifying the interval between optical quality control tests: 4. To change the interval, press and enter a new value followed by.
Cuvette Pathlength Accuracy of total hemoglobin measurements depends on using the correct cuvette pathlength see page 1. The user must check the pathlength whenever using a different lot number of cuvettes. Note: The cuvette lot number and pathlength are included on the carton label and bag for each lot of cuvettes.
Entering a Different Cuvette Pathlength: 1. Press followed by to display the screen for entering the cuvette pathlength: 3. To change the interval, press and enter a new value: Note: Pathlengths ranging from 90 microns to microns can be entered. To change the interval, press and enter a new value. Check that the new pathlength is correctly displayed, then press. Re-Calibration If the ITC AVOXimeter E results are out of range with either blood samples or controls and troubleshooting does not resolve the issue, re-calibration may be necessary.
Press see page 8. The instrument will shut down. Or: 1. Simultaneously press and. It can also calculate differences in oxygen saturation between adjacent anatomical sites Saturation Step-Ups from which blood samples were taken, to aid in diagnosing intracardiac and great vessel shunts.
Enter the age using the number keys, then press. A confirmation prompt is displayed: 4. Check that the age is correctly displayed, then press to confirm. Select for inches or for centimeters , then press. Enter the height using the number keys, then press. Check that the height is correctly displayed, then press to confirm. Select for pounds or for kg , then press.
Enter the weight using the number keys, then press. A confirmation prompt is displayed: Check that the weight is correctly displayed, then press to confirm. Select for male or for female , then press. Alternately, a measured value can be entered. To enter a measured value, delete numerals from the right by pressing once for each numeral to be deleted, then enter the measured value using the number keys.
Check that the correct value is displayed, then press. A confirmation prompt is displayed: 5. Important: In order to calculate saturation step-ups for a patient, the ITC AVOXimeter E results for that patient must be labeled with the Patient ID see page 27 and the anatomical site from which the sample was obtained see page The system searches the database for all results for each main oximetry site for that patient, averages the results for each site, and calculates the oxyhemoglobin saturation step-ups from the average result for each main site.
Note: Use of average readings rather than individual values improves the precision of Saturation Step-Up calculations. The Patient ID for which oxyhemoglobin saturation step-ups will be calculated is displayed: 4. Verify that the Patient ID is correct, then press. Press to display the second page of oxyhemoglobin saturation step-ups for the patient: 6.
Press when done. Calculating Cardiac Output: 1. A menu of flow calculations is displayed: 3. The screen for entering oxygen uptake if not already calculated , venous oxygen content, and arterial oxygen content is displayed: 4. Enter the oxygen uptake if not already calculated , venous oxygen content, and arterial oxygen content, using the number keys. After each value is entered, check that the value is correct and press to move the cursor to the next line.
Note: If oxygen uptake was already calculated, the value will be displayed. If an incorrect value is displayed, delete numerals from the right by pressing once for each numeral to be deleted.
After all values have been entered, press to confirm. The calculated cardiac output is displayed: 6. The menu of flow calculations is again displayed. The screen for entering oxygen uptake, pulmonary venous oxygen content, and pulmonary arterial oxygen content is displayed: 4.
Enter the oxygen uptake, pulmonary venous oxygen content, and pulmonary arterial oxygen content, using the number keys. Note: If an incorrect value is displayed, delete numerals from the right by pressing once for each numeral to be deleted.
The calculated pulmonary flow is displayed: 6. Calculating Flow Ratios from Saturations: 1. The screen for entering arterial and venous oxygen contents is displayed: 5. Enter the arterial oxygen uptake, mixed venous oxygen content, pulmonary venous oxygen content, and pulmonary arterial oxygen content, using the number keys. The calculated flow ratio is displayed: 7. Calculating Flow Ratios from Flows: 1. A menu of flow ratio calculations is displayed: 4.
If needed, enter the pulmonary flow and systemic flow, using the number keys. Note: If pulmonary flow and systemic flow were already calculated, the values will be displayed. Calculating Cardiac Index: 1. The screen for entering cardiac output and body surface area is displayed: 4.
If needed, enter the cardiac output and body surface area, using the number keys. Note: If cardiac output and body surface area were already calculated, the values will be displayed. The calculated cardiac index is displayed: 6. Systemic and Pulmonary Resistances Cardiac output and blood pressure readings are used to calculate systemic resistance and pulmonary resistance. Calculating Systemic Resistance: 1. A menu of resistance calculations is displayed: 3. A screen for entering cardiac output if not already calculated , arterial pressure, and right atrial pressure is displayed: 4.
Enter the cardiac output, arterial pressure, and right atrial pressure, using the number keys. Note: If cardiac output was already calculated, the value will be displayed. The calculated systemic resistance is displayed: 6. The menu of resistance calculations is again displayed. A screen for entering cardiac output if not already calculated , pulmonary artery pressure, and left atrial pressure is displayed: 4.
Enter the cardiac output, pulmonary artery pressure, and left atrial pressure, using the number keys. The calculated pulmonary resistance is displayed: 6. Stroke Index is the stroke volume normalized for body mass. Calculating Stroke Volume: 1. A menu of stroke calculations is displayed: 3. A screen for entering cardiac output if not already calculated and heart rate is displayed: 4. Enter the cardiac output and heart rate, using the number keys. The calculated stroke volume is displayed: 6.
The menu of stroke calculations is again displayed. A screen for stroke volume and body surface area if not already calculated is displayed: 4. If needed, enter the stroke volume and body surface area, using the number keys. Note: If stroke volume and body surface area were already calculated, the values will be displayed. The calculated stroke index is displayed: 6. A message is displayed while the values are printing: 4. The instrument does not respond to keystrokes or cuvette insertion.
Self-Test fails on power-up. The following error message is displayed: "Cuvette Inserted??? The microprocessor is locked up. A cuvette is inserted in the instrument. If a cuvette is not inserted in the instrument, see 2 below: 2. Intensities are out of range. The cable from the detector to the circuit board is disconnected.
There is no power to the LEDs. The LED cable may be loose, disconnected or improperly placed on the connector at one or both ends. One or more LEDs is defective. There is no power to the printer.
The printer is not connected to the instrument. The wrong print mode, baud rate, or parity is selected. Press the and keys simultaneously to turn the instrument off, then turn the instrument back on. Remove cuvette, turn unit off and restart. At "Cuvette Inserted??? Press any key and then press to readjust light sources. When program finishes, turn the instrument off and then on. Open the case and check that the cables are properly connected. Connect the printer to an AC outlet and turn on the printer.
Specify the proper values for the printer see page See above. The optical detector may be contaminated by blood or foreign material. The instrument may require calibration.
The wrong cuvette pathlength is entered into the instrument. Equal values were entered for Pulm. Venous and Pulm. An incorrect entry was made for one of the entered quantities.
A cuvette without a sample was inserted. The cuvette was not properly filled. Run optical quality control. If needed, clean the optical detector see page Re-calibrate the instrument see page Obtain the correct cuvette pathlength value from the cuvette carton and enter the value see page Clean the optical detector see page Obtain correct values for Pulm. Review and, if necessary, re-enter correct values. Test a new cuvette that contains the sample. Ensure that the sample reaches the vent patch when filling and that there are no air bubbles in the optical window.
LED intensities are too low. Remove cuvette, turn unit off, and restart. Verify that the self test passes. The sample is lipemic. If possible, obtain a replacement sample that is not lipemic. THb may actually be low. THb may actually be high. Microprocessor error.
Remove residual dried blood or other foreign matter using a water-dampened cloth. Only qualified personnel should perform maintenance as described in this manual.
Cleaning the Optical Detector Materials Needed: 1. Remove the four screws from the bottom of the instrument, using the 0 Phillips screwdriver. Holding the upper and lower covers together, place the instrument in the upright position, with the keypad to your right.
Slowly and carefully lift the upper cover, keeping it parallel with the lower part of the instrument, until the upper cover is free of the rear panel.
Locate the cable that runs from the battery pack to the main circuit board see below. Disconnect the cable from the J connector on the main circuit board. Tilt the upper cover backward to fully expose the main circuit board and small circuit board on the lower part of the instrument.
Locate the flat keypad cable that runs from the keypad to the main circuit board. Disconnect the cable from the Keyboard J connector on the main circuit board.
Keypad Cable 8. Disconnect the cable from the Display J connector on the main circuit board. Note: Grasp only the connector at the end of the cable. Gently rock the connector in an upward direction to remove it. Locate the flat cable that runs from the black optical unit to the main circuit board. Locate the coaxial cable that runs from the black optical unit to the small circuit board. Disconnect the cable from the J1XX connector on the small circuit board.
The upper cover containing the black optical unit can now be rested on a table with the front panel facing down. Using the 5mm nutdriver, remove the four nuts and washers, if applicable that secure the black optical unit to the front panel. Then remove the black optical unit from the instrument. Take care when separating the two halves of the optical unit.
Do not lose the small torsion spring or the shutter door. Set them aside in a safe place for reassembly. Clean the exposed detector with gauze pads dampened with detergent. Do not use abrasives. Then dry off the detector surface, making sure that it is clean. Remove any other debris inside the optical unit. Reassemble the two halves of the optical unit, making sure that the shutter door and torsion spring are installed correctly the two door ribs will face outward.
One leg of the spring fits into a hole in the door slot, the other leg rests behind the shutter door. Bolt the optical unit back together. Place the optical unit back on the front panel. Using the 5mm nutdriver, secure the black optical unit to the front panel with the four nuts and washers, if applicable.
Position the upper cover containing the black optical unit in an upright position over the lower part of the instrument, with the keypad to the right. Connect the coaxial cable from the optical unit to the J1XX connector on the small circuit board. Connect the flat cable from the optical unit to the J connector on the main circuit board. Connect the flat cable from the keypad to the Keyboard J connector on the main circuit board.
Connect the cable from the battery pack to the J connector on the main circuit board. Slowly and carefully lower the upper cover onto the lower part of the instrument, ensuring that the metal rear panel slides into the center slot at the rear of the upper cover. Note: Ensure that the cables are not snagged or crimped between components when lowering the upper cover onto the lower part of the instrument. Using the 0 Phillips screwdriver, secure the bottom of the instrument to the upper cover with the four screws.
Menu Commands Access Main Menu Press the key when a test is not running or another menu is not active. Page Test Cuvettes Federal, state and local laws and regulations require that hazardous waste be disposed of in a specific manner. Waste material from the AVOXimeter E which may be classified as biohazardous include used cuvettes. It is important that steps be taken to dispose of these materials in accordance with the prevailing regulations in your location.
Page Connections Figure 4. The instrument is factory preset for a standby delay time of 60 minutes, but a time of 10 to minutes can be specified see page Page Optional Materials This eliminates the risk of the instrument powering down during a test.
Fully charged batteries will allow the AVOXimeter E to analyze blood samples continuously for up to 8 hours when the display is set at medium brightness. Press to return to the previous menu, if desired. Setting Up the Instrument The user can specify the display brightness, specify the units that are used for reporting total hemoglobin THb , change the date and time, and specify the length of time that the instrument is idle before it enters the standby mode.
See page 21 for details. The new date is displayed. Press followed by. To delete a user, display that user, press There are 15 main oximetry sites, some of which are divided into optional subsites The instrument starts and performs a series of self-tests.
Do not use samples that contain excessive volumes of anticoagulant or are diluted with saline. Printing all Stored Data If needed Turn on the printer. The last test record is displayed: Press to view additional information, if desired. A confirmation prompt is displayed, indicating that all samples were not printed. Press to return to the previous menu. Page Quality Control Daily optical quality control. Note: If quality control results are out of range, refer to the Troubleshooting section for instructions.
Fill a test cuvette with a liquid control see page Note: Lot numbers of liquid controls can be manually tagged to each liquid QC test as it is run see page Note: Lot numbers of liquid controls must be entered before they can be tagged to a liquid QC test see page Page Calibration Re-Calibration If the AVOXimeter E results are out of range with either blood samples or controls and troubleshooting does not resolve the issue, re-calibration may be necessary.
Please contact Technical Support for more information on re-calibration. The instrument will shut down. Simultaneously press. It can also calculate differences in oxygen saturation between adjacent anatomical sites Saturation Step-Ups from which blood samples were taken, to aid in diagnosing intracardiac and great vessel shunts.
Alternately, a measured value can be entered. Page Saturation Step-Ups Patient ID see page 27 and the anatomical site from which the sample was obtained see page
0コメント